Recall StatusTerminated
Date Initiated2013-12-19
Date Terminated2015-07-07
Product Quantity52
Reason for Recall
3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified
that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw.
Firm Action
URGENT MEDICAL DEVICE RECALL notices were sent to all consignees on December 19, 2013. Distributors who have affected hospitals in their regions were notified and provided with a copy of the notice that went to each hospital. Consignees are being contacted on an individual basis in an attempt to reconcile the product 100%. Fax back forms were provided to all consignees in order for Biomet to track returns. Non-responders will be contacted on an individual basis.
Root Cause
Labeling mix-ups
Distribution Pattern
US Distribution including states of: AZ, CA, GA, IL, KY, MI, NC, NH, NY, OH, PA, SD, TX, UT, and VA.