TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.

FDA Recall Z-0718-03 · Posted 2003-04-10

Recalling FirmKinetic Concepts, Inc
Recall StatusTerminated
Product CodeINX
Date Initiated1997-06-01
Date Terminated2003-07-08
Product Quantity4264

Reason for Recall

Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge.

Firm Action

In approximately November 1997 the firm implemented the TransportAir retrograde for all active units.

Root Cause

Other

Distribution Pattern

Nationwide