Terminated

AxiEM EM Knee Kit, and associated workstation.

FDA Recall Z-0718-06 · Posted 2006-04-07

Recall StatusTerminated
Product CodeHAW
Date Initiated2006-02-01
Date Terminated2006-04-06
Product Quantity84 kits

Reason for Recall

Knee Navigation Disposable Kits were programmed with an incorrect identification number on the tracker ROM chip which would not permit navigation during surgery.

Firm Action

Medtronic field personnel were notified by letter on 02/01/2006 and instructed to immediately contact each affected site and schedule an upgrade with the revised firmware

Root Cause

Other

Distribution Pattern

AZ, CA, CO, GA, IL, LA, MT, NE, NV, OH, TX. No military or government distribution. Foreign distribution to Australia, Austria, Israel, Singapore, Switzerland, UK.