Hamilton Sample Carriers for the ML STAR line (SMP-CAR-24). Product Code: 866.2500; Part Number: 173400. Hamilton Company, Reno, NV 89520 (Accessory on the Microlab STAR series instruments.)

FDA Recall Z-0718-2008 · Posted 2008-02-08

Recalling FirmHamilton Co
Recall StatusTerminated
Product CodeJTC
Date Initiated2007-06-25
Date Terminated2008-04-15
Product Quantity179 units distributed.

Reason for Recall

Incorrectly labeled-- Product has the wrong barcode labeling. The barcode error will not influence the results of testing. The instrument will discover the error and post an error notice.

Firm Action

The firm sent an Urgent: Product Recall letter to consignees on June 25, 2007. The firm also issued an alert via their User news distribution. They instructed consignees to examine their stock and return any affected carriers. A fax back form was supplied with the notification. The firm will be providing replacements.

Root Cause

Error in labeling

Distribution Pattern

Worldwide Distribution --- USA including states of OH, CA, NJ, NC, and countries of Japan, France, Switzerland, Spain, Italy, Germany, and the UK.