Urologix Microwave Delivery System 2.5 - 3.5 For use with prostatic urethra lengths of 2.5-3.5 cm only. Model TA1321D.

FDA Recall Z-0721-2007 · Posted 2007-04-05

Recalling FirmUrologix, Inc.
Recall StatusTerminated
Product CodeMEQ
Date Initiated2007-03-16
Date Terminated2007-05-25
Product Quantity21

Reason for Recall

The catheter in the kit matches the description on the kit and pouch label; however, an incorrect label was placed onto the catheter during manufacturing.

Firm Action

A letter dated 3/16/07 was FedEx overnight to consignees. The letter described the product, problem and gave instructions for returning and replacing the affected product.

Root Cause

Other

Distribution Pattern

Nationwide and the Netherlands.