Barbiturates Intercept Micro-Plate EIA (enzyme immunoassay, barbiturate) Product is intended for use by clinical laboratories in the qualitative determination of barbiturates in oral fluids.

FDA Recall Z-0722-2017 · Posted 2016-12-08

Recall StatusTerminated
Product CodeDIS
Date Initiated2016-11-10
Date Terminated2017-04-17
Product Quantity4 kits

Reason for Recall

Shipping error. Cocaine metabolite Microplate distributed instead of Barbiturate Microplate due to incorrect selection of Microplate during manufacturing.

Firm Action

Due to the small number of customers that were affected by the recall, OraSure notified customers on November 10, 2016, via telephone with a follow-up email. Customers were asked to quarantine the units on hand and they will schedule a pickup by Fed EX to retrieve the affected units. For further questions, please call (610) 882-1820, ext. 1657

Root Cause

Mixed-up of materials/components

Distribution Pattern

Distributed to KS, and VA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.