MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

FDA Recall Z-0722-2024 · Posted 2024-01-12

Recalling FirmCook Medical Llc
Recall StatusOpen, Classified
Product CodeMQG
Date Initiated2023-12-07
Product Quantity14 US; 101 OUS

Reason for Recall

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Firm Action

The firm notified affected customers on December 7, 2023 via URGENT Medical Device Correction letters. Per the letter, customers can continue to use their MINC+ device whilst awaiting the defect correction, but Cook advises customers to be vigilant and monitor for device alerts. If the device produces audible and visual alerts to indicate an error with temperature, customers must immediately move any dishes to another incubator. If no other incubators are available, the device can be reset to normal operation by switching off mains power to the device for ten seconds then turning the power back on. The software/firmware update will undergo verification testing including compliance testing prior to release. The updated software/firmware is expected to be available on 01 February 2024. An authorized service agent will contact customers to arrange for the device to be corrected at their facility. The service agent will be updating the software/firmware of the device.

Root Cause

Device Design

Distribution Pattern

Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM