Intuiitve Surgical da Vinci 8 mm EndoWrist Curved Scissors, in box. All lots. P/N 400178-01

FDA Recall Z-0723-05 · Posted 2005-04-22

Recall StatusTerminated
Product CodeLRW
Date Initiated2005-03-09
Date Terminated2005-05-13
Product Quantity278 units have been distributed.

Reason for Recall

Blades on the scissor may break and separate from the main unit as a result of corrosion damage.

Firm Action

The firm has sent notification letters to its consignees on 03/09/2005. Consignees are requested to return the feedback form provided by Intuitive Surgical, via Federal Express.

Root Cause

Other

Distribution Pattern

The product was distributed to 84 medical facilities, located throughout the US and Switzerland, Belgium, Germany, Australia, Turkey and France.