Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (case containing 20 boxes), Manufactured by Hamilton Company, Reno, Nevada
FDA Recall Z-0723-2007 · Posted 2007-04-07
Advertisement
Recalling FirmHamilton Co
Recall StatusTerminated
Product Code-JTC
Date Initiated2007-01-30
Date Terminated2007-11-28
Product Quantity37 cases
Reason for Recall
Some of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of causing a missed pipetting.
Firm Action
The firm initiated its recall on January 30, 2007. The initial notification was sent via certified mail and e-mail. Subrecalls will be conducted by the distributor.
Root Cause
Other
Distribution Pattern
Nationwide (via a single distributor in New Jersey)
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.