Open, Classified

IntelePACS - InteleConnect / TechPortal

FDA Recall Z-0723-2026 · Posted 2025-11-26

Recall StatusOpen, Classified
Product CodeLLZ
Date Initiated2025-10-17
Product Quantity19

Reason for Recall

Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.

Firm Action

On 10/17/2025, correction notices were emailed to customers who were asked to do the following: 1. Schedule Feature Release Upgrade (FRU) update immediately with the firm's Customer Success Manager (CSM). 2. Interim Mitigation: Until the update is applied, please advise all staff to verify the 'Relevant Clinical Info' field after any Case Editor modifications and consider temporarily documenting critical clinical information in alternative fields as backup. 3. complete and return the Acknowledgement Form via email to [email protected]. For more information or to ask any questions about the notification contact the firm's regulatory affairs department at [email protected]

Root Cause

Software Design Change

Distribution Pattern

US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑