IMMULITE 2000 Intact PTH kit

FDA Recall Z-0724-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeCEW
Date Initiated2004-02-23
Date Terminated2004-12-02
Product Quantity2416 kits

Reason for Recall

High bias noted in samples collected with EDTA. Indication is to not run samples with EDTA.

Firm Action

Recall was by letter mailed / faxed on February 23, 2004. Users are instructed to remove EDTA plasma samples for consideration for testing and correct instructions in the package insert.

Root Cause

Other

Distribution Pattern

North America, Central America, South America, Europe, Asia, South Africa, Middle East. Domestic includes: VA Hospital, St. Louis MO VA Medical Center, Denver, CO VA Medical Center, Richmond, VA VA Medical Center, West Haven, CT Walter Reed Army Medical Center, Washington DC