Visions PV .035 Digital IVUS Catheter, Catalog Number/Part Number 88901/400-0200.173 Product Usage: The Volcano PV .035 Digital IVUS Catheter is an over the wire imaging catheter with a digital ultrasound transducer on the distal tip. The transducer utilizes a 64-element cylindrical array that radiates acoustic energy into the surrounding tissue and detects the subsequent echoes. The information from the echoes is used to generate real-time images of the peripheral vessels. The devices are provided sterile and are single use disposable devices.
FDA Recall Z-0726-2019 · Posted None
Advertisement
Recalling FirmVolcano Corporation
Recall StatusTerminated
Product CodeOBJ
Date Initiated2018-10-17
Date Terminated2021-04-30
Product Quantity6870 units
Reason for Recall
Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the guidewire through the y-port of the catheter.
Firm Action
The firm disseminated a Medical Device Advisory Notice by mail on 10/17/2018. The letter requested that the consignee notify the complaint department at 800-228-4728 Option 3 or email [email protected] to report an ability to insert the guidewire into the catheter. The defective units will be replaced.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.