ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system

FDA Recall Z-0727-2021 · Posted 2021-01-28

Recall StatusTerminated
Product CodeOWB
Date Initiated2020-12-08
Date Terminated2024-06-21
Product Quantity91

Reason for Recall

After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Nationwide