Centurion Healthcare Products Laceration Tray, sterile; (Reorder SUT8805A) Tri-State Hospital Supply Co., Howell, MI. 48844. The primary kit is labeled as containing 1 needle, 21G x 1-1/2", among other components, however some kits may contain a 21Gx1" needle. The label on the Kendall Monoject Magellan 21G Safety Needle itself is the correct size.

FDA Recall Z-0730-2008 · Posted 2008-02-23

Recall StatusTerminated
Product CodeKDC
Date Initiated2007-10-26
Date Terminated2008-03-13
Product Quantity396 kits.

Reason for Recall

Incorrect Needle size - Some of the kits contain a 1 inch 21G needle instead of the 1-1/2 inch 21 needle specified on the kit label.

Firm Action

An Urgent Product Recall Notice was sent on October 25, 2007 to Distributors, Distributors customers, and all third party consignee. The letter provided information concerning the product recall, and requested removal of all product inventory. A product return form requesting information on products still in the facility was enclosed. A Tri-State Representative will be in contact regarding retrieval of the product.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Nationwide Distribution, including states of Massachusetts and Washington.