Palacos R+G Radioplaque Bone Cement 1 x 40 g Single (with Gentamicin), 40.8 g Methyl Acrylate Copolymer, 20 ml Methyl Methacrylate Monomer 0.5 g Gentamicin, Distributed by Zimmer Dover, OH 44522 The cement is indicated for use in the second stage of a two stage revision for total joint arthroplasty after the initial infection has been cleared.

FDA Recall Z-0730-2014 · Posted 2014-01-13

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeMBB
Date Initiated2013-11-04
Date Terminated2014-10-08
Product Quantity1,534 units

Reason for Recall

The patient label associated with the product may be incorrect in that the lot number on the patient label may be different than the lot number on the outer carton label.

Firm Action

Zimmer sent an Urgent Device Removal letter dated November 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification, idetntify the product and remove from use. If the affected product has been utilized, follow institutional protocol to determine if it is appropriate to place a copy of the notification letter into patient records for the affected product. Complete the attached Response Form and return it via fax to 800-871-7270. Customers were instructed to return a copy of the completed response form along with their returned product to ensure proper credit. Customers with questions were instructed to call 866-608-3941. For questions regarding this recall call 800-613-6131.

Root Cause

Other

Distribution Pattern

Nationwide Distribution including AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI, WV & WY.