iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101
FDA Recall Z-0731-2019 · Posted None
Advertisement
Recalling FirmConformis, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2018-10-16
Date Terminated2020-06-29
Product Quantity1
Reason for Recall
The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.
Firm Action
The customer was contacted via email on 10/16/18.
Root Cause
Under Investigation by firm
Distribution Pattern
The products were distributed to the following US states: TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.