Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

FDA Recall Z-0731-2019 · Posted None

Recalling FirmConformis, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2018-10-16
Date Terminated2020-06-29
Product Quantity1

Reason for Recall

The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.

Firm Action

The customer was contacted via email on 10/16/18.

Root Cause

Under Investigation by firm

Distribution Pattern

The products were distributed to the following US states: TN.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑