NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).
FDA Recall Z-0735-04 · Posted 2004-07-20
Advertisement
Recalling FirmBiomerieux, Inc.
Recall StatusTerminated
Product CodeJJQ
Date Initiated2004-02-26
Date Terminated2004-03-24
Product Quantity3 units
Reason for Recall
Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Firm Action
This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
Root Cause
Other
Distribution Pattern
CT, MD, MT