Terminated

Roche COBAS Integra DIG, Digoxin Reagent for use on the COBAS INTEGRA models 400, 400 plus, 700 and 800 analyzers, as well as, Cobas c 501 analyzer; Catalog Number 20737836322.

FDA Recall Z-0736-2007 · Posted 2007-04-13

Recall StatusTerminated
Product CodeKXT
Date Initiated2007-03-16
Date Terminated2008-06-12
Product Quantity4,735

Reason for Recall

The lower detection limit (LDL) may be a higher value than stated in the labeling. Note: To resolve this issue, the Digoxin LDL claim has been revised to 0.3 ng/mL for all reagent lots.

Firm Action

A recall letter dated 3/16/07 was sent to each consignee advising them that the LDL claim in the label has been revised to 0.3 ng/ml for all reagent lots.

Root Cause

Other

Distribution Pattern

Nationwide.