CoolSeal Generator, REF: CSL-200-50

FDA Recall Z-0736-2022 · Posted None

Recalling FirmBolder Surgical, LLC
Recall StatusTerminated
Product CodeGEI
Date Initiated2022-01-25
Date Terminated2024-05-15
Product Quantity10

Reason for Recall

A potential manufacturing/assembly defect could result in the generator not meeting IEC 60601 standards for medical electrical equipment, and it can potentially, under specific conditions, lead to the delivery of unintended energy to the patient.

Firm Action

On 01/25/22, recall notices were hand delivered to U.S. customers and emailed to O.U.S. customers. Customers were asked to do the following: Discontinue use and return affected devices to the recalling firm, and complete and return the acknowledgment form. Customers with additional questions were encouraged to contact the recalling firm at 1-866-683-1743.

Root Cause

Process control

Distribution Pattern

US: NY, MO, DE, PA, FL OUS: Italy, Chile, UK