Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001

FDA Recall Z-0737-2024 · Posted 2024-01-17

Recalling FirmSteris Corporation
Recall StatusOpen, Classified
Product CodeLRW
Date Initiated2023-11-16
Product Quantity4,214 units

Reason for Recall

Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.

Firm Action

Steris issued Urgent Medical Device Recall Letter to Distributors and Customers on November 16, 2023. Letter states reason for recall, health risk and action to take: Please ensure the following steps are completed: 1. Please immediately inspect your on-hand inventory for product affected by this recall. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. STERIS will coordinate credit issuance upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: [email protected] or via fax to (440) 392-8963. If you have any questions regarding this recall, please contact [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution.

Official Source

View FDA Recall →