Terminated

Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA

FDA Recall Z-0739-04 · Posted 2004-07-20

Recalling FirmAsahi Medical Co Ltd
Recall StatusTerminated
Product CodeKDI
Date Initiated2001-10-01
Date Terminated2004-03-24
Product Quantity180,000 units estimated

Reason for Recall

Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.

Firm Action

Asahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.

Root Cause

Other

Distribution Pattern

New Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →