Terminated

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

FDA Recall Z-0741-2013 · Posted 2013-01-31

Recalling FirmHamilton Medical Inc
Recall StatusTerminated
Product CodeCBK
Date Initiated2012-12-19
Date Terminated2014-04-22
Product Quantity97 total ventilators; 69 of part number 161005 and 28 of part number 161006

Reason for Recall

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

Firm Action

Hamilton sent a Field Safety alert letter dated January 11, 2013 to all customers. The letter identified the affected product, explained the potential problem and a included Medical Device Safety Alert and Corrective Action document. Hamilton is working on a new software version for this issue. Customers are asked to update their operator manual with the update oxygen consumption formula page which provided revised information for calculating oxygen capacity. When the new software is available, Hamilton will provide loaner devices to each location while their unit is being upgraded. Questions should be addressed to Hamilton Medical at 800-426-6331 ext 215

Root Cause

Software design

Distribution Pattern

US Nationwide Distribution including the country of Switzerland.