Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024
FDA Recall Z-0741-2022 · Posted None
Advertisement
Recalling FirmCytocell Ltd.
Recall StatusTerminated
Product CodeQDI
Date Initiated2022-01-18
Date Terminated2024-11-08
Product Quantity19 kits
Reason for Recall
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Firm Action
Cytocell issued Urgent Medical Recall Reference: VC/2022/001 letter on 1/18/22 to Distributors and End Users via email. Letter states reason for recall, health risk and action to take:
Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of the probe being manufactured with the fluorophores the incorrect way around.
Please complete and return the enclosed response form as soon as possible.
If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or [email protected].
Root Cause
Software Manufacturing/Software Deployment
Distribution Pattern
US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.