Terminated

Alfa Wassermann System Diluent 750 mL Manufactured for Alfa Wassermann Diagnostic Technologies LLC 4 Henderson Drive West Caldwell, NJ 07006 Alfa Wassermann B.V. Pompmolenlaan 24 3447 GK, Woerden The Netherlands Common name is water solution. Intended to dilute reagents and rinse the parts of a clinical chemistry instrument.

FDA Recall Z-0743-2013 · Posted 2013-01-28

Recalling FirmAlfa Wassermann, Inc.
Recall StatusTerminated
Product CodeCEO
Date Initiated2012-10-10
Date Terminated2013-10-25
Product Quantity4,104 kits

Reason for Recall

Inorganic phosphorus assays using System Diluent Lot 155599 may not product acceptable calibration results.

Firm Action

Alfa Wassermann sent an important advisory notice/product response form to their distributors on October 10, 2012. The letter identified the product, the problem, and the action to be taken by the customer. The letter informs customers about the affected lot and provide instruction regarding the receipt of replacement product. Any recalled product should be destroyed. If there is further distribution, those customers should provide Alfa with a distribution list so that they may be notified. Any questions can be directed to (866) 419 - 2532.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the countries of Canada and Cayman Islands

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.