Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8161. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.
FDA Recall Z-0744-2012 · Posted 2012-01-13
Advertisement
Recalling FirmBaxter Healthcare Corp.
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-07-14
Date Terminated2012-01-13
Product Quantity2 units
Reason for Recall
Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.
Firm Action
Baxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.
Root Cause
Employee error
Distribution Pattern
Nationwide distribution: Puerto Rico.
Related FDA 510(k) Records