ACIST Medical Systems, BT2000 Automated Manifold Kit (SKU 014613) and CPT2000 (SKU 014623) Automated Manifold Kits, Sterile R, Rx Only, Made in Mexico. The manifold body is identical in the BT2000 and CPT2000 kits. The BT2000 and CPT2000 kits facilitate connection to, and delivery of, either contrast or saline into the patient via a coronary catheter (not provided by ACIST). To be used for the controlled infusion of radiopaque contrast media for angiographic procedures

FDA Recall Z-0744-2016 · Posted 2016-02-02

Recalling FirmAcist Medical Systems
Recall StatusTerminated
Product CodeDXT
Date Initiated2015-12-21
Date Terminated2016-07-22
Product Quantity48,614 cartons or 486,140 units

Reason for Recall

Reports related to ingress of air into the manifold.

Firm Action

Acist sent an Urgent Medical Device Recall letter dated December 21, 2015, via e-mail to all affected consignees. The letter was addressed to Distribution Partner or Cardiac Cath Lab / Material Manager in the USA. In Europe, the letter was addresses to Purchasing Department. The letter described the problem and the product involved in the recall. Advised consignees to return affected units and to complete and return the Medical Device Recall Form. Distributors were further advised to send the recall letter to their customers. For questions or concerns, contact in US, ACIST Customer Service 1-888-670-7701 or your ACIST Sales Representative. Europe Direct contact ACIST Europe Customer Service +44 (0) 845 548 12 80 or contact your ACIST Sales Representative, Europe Indirect contact ACIST Europe Customer Service +31 45 750 70 25 or contact your Distributor. Distributors contact: ACIST Europe BV, Office: +31 (0)45 750 7015, Mobile: +31 (0)6 233 648 46.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US including all states except DE, MT, RI, UT, & VT and Internationally to Belgium, France, Germany, Iceland, Ireland, Luxembourg, Netherlands, Reunion, United Kingdom, Australia, Austria, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Estonia, Greece, Hong Kong, Hungary, Islamic Republic of Iran, Israel, Italy, Japan, Republic of Korea, Kuwait, Latvia, Lithuania, Mexico, New Caledonia, Norway, Panama, Poland, Portugal, Saudi Arabia, Serbia, Slovenia, Bosnia and Herzegovina, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and Venezuela.