Terminated

DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621

FDA Recall Z-0745-2016 · Posted 2016-02-02

Recall StatusTerminated
Product CodeIYE
Date Initiated2016-01-08
Date Terminated2016-10-05
Product Quantity3

Reason for Recall

In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.

Firm Action

Philips sent an "Urgent Medical Device Correction" letter dated 16-DEC-2015 to affected customers. The letter described the Affected Products, Problem Description, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer / User, Actions Planned by Philips and Further Information and Support (contact Philips Customer Care Solutions Center at 1-800-722-9377 ).

Root Cause

Device Design

Distribution Pattern

US: Nationwide Distribution in the states of FL, OH, and MO.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.