Terminated

MAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Injection System, or Medrad Mark V ProVis Injection System, Catalog Numbers: Catalog numbers: OCS CEIL (58-S, 58-P, 85-S, 85-P, 100-S, 100-P, 70Si, 97Si), OCS TRACK (58-S, 58-P, 80-S, 80-P), OCS WALL (S, P), OCS (115A, 115CT, 125CT, 125A, 135A, 135CT, 145, 155A, 165A, 115AS, 115CTS), EOC (700, 700S), EOW 700, EOA 700

FDA Recall Z-0745-2019 · Posted None

Recall StatusTerminated
Product CodeDXT
Date Initiated2018-12-14
Date Terminated2020-06-16
Product Quantity16656

Reason for Recall

If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.

Firm Action

Urgent Medical Device Field Corrective Action notification letters were sent to customers on 12/12/18.

Root Cause

Device Design

Distribution Pattern

The products were distributed US nationwide. The products were distributed to the following foreign countries: Algeria, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Cote D'Ivoire, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Greenland, Guam, Vatican City State, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mauritius, Mayotte, Mexico, Monaco, Mongolia, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Reunion, Russian Federation, Saint Pierre and Miquelon, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey UAE, UK, Uruguay, Venezuela, Vietnam.