Terminated

PERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.

FDA Recall Z-0747-2019 · Posted None

Recalling FirmEthicon, Inc.
Recall StatusTerminated
Product CodeGAP
Date Initiated2018-12-18
Date Terminated2020-05-29
Product Quantity840

Reason for Recall

The sutures inside the packaging are not the same size or type as indicated on the label.

Firm Action

Ethicon sent an Urgent Medical Device Recall notification letters to customers on 12/18/18. The letter identified the affected product, problem and actions to be taken. For questions contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266).

Root Cause

Packaging process control

Distribution Pattern

The products were distributed to the following US states: AR, CA, FL, GA, KS, LA, MO, MS, ND, NE, NJ, NY, OH, PA, and WI.