ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235

FDA Recall Z-0747-2022 · Posted None

Recall StatusOpen, Classified
Product CodeLHN
Date Initiated2022-02-15
Product Quantity15 units US; 22 units OUS

Reason for Recall

When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range

Firm Action

IBA issued Urgent Medical Device Correction notification on February 15, 2022 directly to customer either by hand or by email. Letter states reason for recall, health risk and action to take: For Proton Therapy sites where the system is operated by the customer: - Warn operators to not manually change the beam range (action called manual set range ) while a patient is on the Patient Positioning System. - IBA will provide a training to all operators about the correct use of the action called manual set range and the risk associated to its use while a patient is on the Patient Positioning System. This training will be provided on your site by August 31, 2022. Contact: [email protected]. Help Desk: +32 2 507 20 81 (available 24/7)

Root Cause

Device Design

Distribution Pattern

IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,