Mobilett XP, mobile X-ray system, Model number 1818447
FDA Recall Z-0750-2007 · Posted 2007-04-19
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeIZL
Date Initiated2007-03-21
Date Terminated2007-08-07
Product Quantity36 units
Reason for Recall
Mounting bolts for tank fork assembly and collimator flange may become loose.
Firm Action
Siemens Service Personnel are visiting each consignee, beginning March 21, 2007, to replace the required mounting bolts and tighten them according to required torque specifications.
Root Cause
Other
Distribution Pattern
The products were shipped to medical facilities nationwide.