Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002 Software Version: N/A Product Description: Locking Cap Component: No

FDA Recall Z-0751-2025 · Posted 2024-12-23

Recall StatusOpen, Classified
Product CodeNKB
Date Initiated2024-10-29
Product Quantity92 devices

Reason for Recall

Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.

Firm Action

Update 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affected part(s) using the provided label to return the inventory in their possession. Any questions contact Dalton Daggs at [email protected] or Danielle Nelson-Archer at [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

U.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑