METRx II Tube, 22mm x 9cm stainless steel tubular retractor, Part No. 9569736

FDA Recall Z-0753-06 · Posted 2006-04-12

Recall StatusTerminated
Product CodeLXH
Date Initiated2006-01-25
Date Terminated2007-07-05
Product Quantity1

Reason for Recall

Due to a manufacturing error after autoclaving the Tubular Retractor coating, Physical Vapor Deposition (PVD) black Titanium Nitride (TiN), may rub off.

Firm Action

The firm sent out a recall letter dated January 25, 2006 stating the problem and requesting the return of all unused product.

Root Cause

Other

Distribution Pattern

IL, IN, MI, CA, Japan, Netherlands