Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

FDA Recall Z-0753-2024 · Posted 2024-01-18

Recall StatusOpen, Classified
Product CodeJIX
Date Initiated2023-11-20
Product Quantity108 kits

Reason for Recall

Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

Firm Action

Randox issued Urgent Medical Device Correction letter on 11/20/23 to the distribution center within the USA. The US distributor contacted the customer direct via email. Letter states reason for recall, health risk and action to take: Review your calibrator inventory of this lot and assess your laboratories needs for reimbursement for discarded inventory. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form, 12187-QA to [email protected] within five working days. If you have any questions or concerns please contact Randox Technical Services.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide including Puerto Rico.