Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351
FDA Recall Z-0753-2024 · Posted 2024-01-18
Advertisement
Recalling FirmRandox Laboratories Ltd.
Recall StatusOpen, Classified
Product CodeJIX
Date Initiated2023-11-20
Product Quantity108 kits
Reason for Recall
Target for Alkaline Phosphatase (ALP) for the AMP
optimized to IFCC 37C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
Firm Action
Randox issued Urgent Medical Device Correction letter on 11/20/23 to the distribution center within the USA. The US distributor contacted the customer direct via email.
Letter states reason for recall, health risk and action to take:
Review your calibrator inventory of this lot and assess your laboratories needs for
reimbursement for discarded inventory.
" Discuss the contents of this notice with your Medical Director.
" Complete and return the response form, 12187-QA to [email protected]
within five working days.
If you have any questions or concerns please contact Randox Technical
Services.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide including Puerto Rico.