Terminated

Percuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.

FDA Recall Z-0754-2019 · Posted None

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeGEI
Date Initiated2018-01-03
Date Terminated2020-10-30
Product Quantity1962 units

Reason for Recall

There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.

Firm Action

The firm, Teleflex, sent an "Urgent Medical Device Recall" letter dated 1/3/2019 to its customers on 01/03/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1.If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected] or hand it to your sales representative. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-396-2111.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Iran, Ireland, Italy, Japan, Lebanon, Netherlands, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, and the United Arab Emirates.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →