Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS 3600, VITROS 5600, and VITROS ECi/ECiQ Systems.
FDA Recall Z-0755-2013 · Posted 2013-01-30
Recalling FirmOrtho-Clinical Diagnostics
Recall StatusTerminated
Product CodeJJE
Date Initiated2012-12-13
Date Terminated2018-07-05
Product QuantityDomestic: 15,891; Foreign: 5237
Reason for Recall
Some VITROS Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS System instead of the intended VITROS Vapor Adsorption Cartridge which may bias results.
Firm Action
Ortho-Clinical Diagnostics sent URGENT PRODUCT CORRECTION Letters (dated 12/13/12) along with Q&A and Confirmation of Receipt Forms to distributors and consignees on 12/13/12. They were advised of the issue and instructed to discard their current inventory of VITROS Vapor Adsorption Cartridges. On 12/13/12, foreign affiliates were informed by e-mail of the issue and instructed to notify their consignees of the issue and actions.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.