Olympus TJF-Q180V flexible gastrointestinal Duodenoscope Product Usage: This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

FDA Recall Z-0757-2016 · Posted 2016-02-04

Recall StatusTerminated
Product CodeFDT
Date Initiated2016-01-08
Date Terminated2017-01-09
Product Quantity4436

Reason for Recall

Olympus America Inc is conducting a voluntary removal/corrective action of all TJF-Q180V duodenoscopes in order to replace the forceps elevator mechanism Olympus is replacing the forceps elevator mechanism with a new forceps elevator design consistent with the design specification in the recently cleared TJF-Q180V 510k.

Firm Action

Olympus America Inc. sent an Urgent Notification letter dated January 15, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory for the affected product, Olympus will contact facilities to make arrangements for return of the TJF-Q180V duodenoscope(s) for the elevator mechanism replacement. For questions call (484) 896-5688.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide Distribution except PR