Stryker Reflex Hybrid Screw Extractor; Not Sterile: Manufactured by: Stryker Spine SAS, Cestas, France; Distributed in the USA by: Stryker Spine, 2 Pearl Court, Allendale, NJ. The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.

FDA Recall Z-0762-2010 · Posted 2010-04-12

Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeKWQ
Date Initiated2009-10-12
Date Terminated2012-09-06
Product Quantity1,791

Reason for Recall

The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.

Firm Action

An "Urgent Product Recall" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009. The recall letter described the affected products, issue and potential hazards. Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter. Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370. Direct questions about the recall to Stryker Spine by calling 1-201-760-8206.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →