iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.
FDA Recall Z-0762-2020 · Posted None
Advertisement
Recalling FirmDeerfield Imaging, Inc.
Recall StatusTerminated
Product CodeLNH
Date Initiated2019-11-26
Date Terminated2020-08-04
Product Quantity6
Reason for Recall
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible
Firm Action
An Urgent Field Safety Notice letter was sent to customers on 11/26/2019 via FedEx.. The letter state the issue, potential risk, the actions to be taken, and stated that the device will be modified. Questions can be directed to the service line at 1-866-475-0525.
Root Cause
Component design/selection
Distribution Pattern
CT, MA, VA. International distribution: Australia, Canada, and Qatar.