Recall StatusTerminated
Date Initiated2015-11-03
Date Terminated2016-08-19
Product Quantity2 subject to correction and removal
Reason for Recall
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Firm Action
The letters for the customers in the United States were sent on 11/3/2015, via certified return receipt and were addressed to the Risk Manager.
Philips will perform a test on the 15 affected Allura Xper R8.2 FD 20C systems to see if the motion controller amplifier of the brake motor works correctly. If the test will not pass, the motor will be replaced.
The Recall Package is sent to all Key Markets worldwide. The Key Market of an affected country will contact the facility where these systems were installed.
Distribution Pattern
Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.