Terminated

MicroSTAAR Injector w/ Foam Tip Plunger, Model MSI-PF

FDA Recall Z-0766-04 · Posted 2004-07-20

Recalling FirmStaar Surgical Co Inc
Recall StatusTerminated
Product CodeKYB
Date Initiated2004-02-23
Date Terminated2005-01-04
Product Quantity690 injectors

Reason for Recall

Injector mold design/manufacture and wear caused material flash to develop preventing assembly of the device prior to use.

Firm Action

Firm will be sending thier representatives to customers to inspect inventory and arrange return of product.

Root Cause

Other

Distribution Pattern

Nationwide.

Official Source

View FDA Recall →