Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Miltex Stainless Steel Disposable Scalpel #22, Part Number 4-422. Each box contains 10 individually packaged scalpels.

FDA Recall Z-0767-2007 · Posted 2007-04-21

Recalling FirmMiltex, Inc.
Recall StatusTerminated
Product CodeGES
Date Initiated2007-04-09
Date Terminated2007-08-07
Product Quantity400 boxes

Reason for Recall

Sterility may be compromised based on incomplete package seals (manufacturer notified Miltex)

Firm Action

The recalling firm issued a recall letter dated 4/9/07 informing the customers of the problem and the need to return the product. The letter was sent via UPS with a return receipt required to verify delivery. Distributors were notified to sub-recall.

Root Cause

Other

Distribution Pattern

Nationwide to distributors in CA, FL, IA, ID, NM, NJ, OH, PA, RI, and VA.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑