Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications
FDA Recall Z-0767-2013 · Posted 2013-02-01
Recalling FirmPhilips Healthcare Inc.
Recall StatusTerminated
Product CodeKPR
Date Initiated2013-01-16
Date Terminated2017-01-23
Product Quantity14 units
Reason for Recall
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Firm Action
Phillips issued an Urgent Field Safety Notice to all customers on January 14, 2013, to notify users that the Phillips EasyDiagnost Eleva systems, release 3.0 has a mirror icon that could be misinterpereted and could potentially lead to misdiagnosis. Philips will upgrade the system to version 3.1 which involves software and hardware upgrades of the system. This software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Additionally, Philips will provide updated Instructions for Use (IFU) for the Eleva workspot and the EasyDiagnost 3.1 system and an addendum to the EasyDiagnost system IFU. Users will be contacted by Phillips to schedule the installation. For questions contact your local Philips representative: 1-800-722-9377.
Root Cause
Software design (manufacturing process)
Distribution Pattern
US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.