MEVION S250 Radiation therapy system
FDA Recall Z-0767-2016 · Posted 2016-02-08
Advertisement
Recalling FirmMevion Medical Systems, Inc.
Recall StatusTerminated
Product CodeLHN
Date Initiated2015-11-18
Date Terminated2017-04-03
Product Quantity4
Reason for Recall
The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.
Firm Action
Customer notification letter was sent via certified mail on November 18, 2015. The notification explains the issue and actions to be taken by the customer.
Root Cause
Labeling design
Distribution Pattern
US nationwide distribution including FL, NJ, OK, and MO.