Medfusion 3500 Syringe Infusion Pump with PharmGuard Medication safety Software versions 4.1.3 and 4.1.4. Medex, 4350 Green Parkway, Suite 200. Duluth, GA 30096
FDA Recall Z-0768-2007 · Posted 2007-04-26
Advertisement
Recalling FirmSmiths Medical MD, Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2007-02-03
Date Terminated2011-02-10
Product Quantity23,519
Reason for Recall
Software anomaly that causes the pump to experience a sudden and unexpected failure. This anomaly occurs during the pump's startup sequence when the pump's Preventative Maintenance Alarm goes off, preventing the pump from completing its power up routine. Lock-up occurs and the pump will not function.
Firm Action
Consignees were initially notified of the recall via a Customer Information Bulletin which was sent via electronic mail 2/3/07 and confimed with a telephone conversation. A Product Recall Notification and a Product Recall Bulletin will be sent to all consignees via the mail on 2/16/07.
Root Cause
Other
Distribution Pattern
Nationwide, Canada, and Italy.