Terminated

Li-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400. Also branded as Reliability Plus byDrive. The Portable Oxygen Concentrator is intended to deliver concentrated oxygen for adult patients with chronic pulmonary diseases.

FDA Recall Z-0768-2013 · Posted 2013-02-02

Recalling FirmOxus America, Inc.
Recall StatusTerminated
Product CodeCAW
Date Initiated2013-01-14
Date Terminated2013-07-17
Product Quantity1724 total units

Reason for Recall

Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.

Firm Action

Oxus, Inc. sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 11, 2013 to all affected customers.The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-888-475-1568 for information related to this recall.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution-USA including the states of New York, Michigan, Colorado, Texas, Iowa, Florida, Georgia, Ohio, Wisconsin, Oregon, Alabama, Illinois, Utah, Oklahoma, West Virginia and California and the countries of Canada, European Union, South Africa, Korea and Brazil.