MiniCap Extended Life PD Transfer Set
FDA Recall Z-0768-2021 · Posted None
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusTerminated
Product CodeKDJ
Date Initiated2020-11-10
Date Terminated2023-03-24
Product Quantity72,771 units
Reason for Recall
Potential for no-flow and leaks under the twist clamp.
Firm Action
Urgent Medical Device Recall FA-2020-055 was sent to Peritoneal Dialysis Centers via U.S.P.S., first class mail on November 10, 2020. Urgent Medical Device Recall FA-2020-055 was sent via U.S.P.S., first class mail on November 13, 2020 to Home Patients who received impacted lots directly from Baxter. Out of an abundance of caution, Baxter will also be mailing the communication to all home patients who receive treatment at Peritoneal Dialysis Centers which received the impacted lots of transfer sets, regardless if the patient received the set or not.
Root Cause
Process change control
Distribution Pattern
Worldwide and US (Nationwide), Canada, Brazil, Mexico, Bermuda, and Indonesia
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.