Remel Blood Agar/MacConkey Agar Bi-Plate, REF: R02049
FDA Recall Z-0769-2021 · Posted None
Advertisement
Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeJSI
Date Initiated2020-12-10
Date Terminated2023-07-06
Product Quantity637 units
Reason for Recall
The blood agar side of the bi-plate may not grow some patient isolates or QC organisms as intended.
Firm Action
The recalling firm issued a recall notification letter dated 12/16/2020 to all impacted customers. The letter states that all impacted product has since expired on 12/10/2020. Customers are asked to review of reported test results should be determined by the appropriate technical expert. Customers are also asked to keep the notification letter on file. The notice should be passed on to all who need to be aware within the organization or to any organization where the potentially affected products have been transferred. Questions may be directed to the Technical Services Department at 1-800-255-6730 (US) or via email: [email protected].
Root Cause
Mixed-up of materials/components
Distribution Pattern
Distribution in United States - AL, FL, GA, LA MS, NC, SC, TN