HiSpeed FX/I (2200997 with Gantry mo. #2200997)

FDA Recall Z-0771-2011 · Posted 2011-01-04

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeJAK
Date Initiated2010-02-01
Date Terminated2014-05-21
Product Quantity102 total

Reason for Recall

A defect with the hardware and software displaying the incorrect dose data.

Firm Action

Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Nationwide Distribution