NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.

FDA Recall Z-0771-2020 · Posted None

Recalling FirmNidek, Inc.
Recall StatusTerminated
Product CodeNQE
Date Initiated2019-02-25
Date Terminated2021-06-15
Product Quantity13 devices

Reason for Recall

The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

Firm Action

The recalling firm issued letters dated 2/7/2019 via FedEx explaining the issue, informing the consignee there was no risk associated with use of the device, and that the firm is preparing for the replacement parts in order to schedule for the correction to be made.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.